Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 126

Advertising
background image

IM 12B07D02-01E

12-14 Appendix

2/3

QIS 12B07D02-71E

Table 1

Reference Temperature

(°C)

Resistance Box Resistance

(:)

Data Display

(°C)

-20 921.6

-20.0

±0.3

+25 1097.3

+25.0

±0.3

+130 1498.3

+130.0

±0.3

3.4 pH

Indication

Check

Set the instrument as follows.

Decade resistance box:

1000 Ω

Change the simulation input value of the standard voltage generator as shown in Table 2. The

corresponding pH data must be within the range.

Table 2

Check Point (pH)

Simulation Input (mV)

Data Display (pH)

0 +414.1

0.00

±0.01

7 0.0

7.00

±0.01

14 -414.1

14.00

±0.01

3.5 ORP

Indication

Check

Put the instrument in ORP mode by following the steps below.

a. Press the [ ] key. The message display will show “ SERV.”

b. Press the [YES] key. The message display will show “ CODE” and the data display will

show “00” with the first digit of 0 flashing.

c. Press the [>] key once. The second digit of 0 flashes.

d. Press the [

š

] key once. The data display will show “01” with 1 flashing.

e. Press the [ENT] key. The message display will show PH.ORP.”

f. Press

the

[

š

] key once. The data display will change from “0” to “1.”

g. Press the [ENT] key.

h. Turn on power again.

In this state, change the simulation input of the standard voltage generator as shown in Table 3.

The corresponding data must be within the range.

Table 3

Simulation Input (mV)

Data Display (mV)

-1500

-1500

±1

0 0

±1

+1500

+1500

±1

Return to PH mode by following the steps below and then the tests are completed.

a. Press the [ ] key. The message display will show “ SERV.”

b. Press the [YES] key. The message display will show “ CODE” and the data display will

show “00” with the first digit of 0 flashing.

c. Press the [>] key once. The second digit of 0 flashes.

d. Press the [

š

] key once. The data display will show “01” with 1 flashing.

e. Press the [ENT] key. The message display will show PH.ORP.”

f. Press

the

[

š

] key once. The data display will change from “1” to “0.”

g. Press the [ENT] key.