Specification 2-17

x

In each I.S. Fieldbus segment only one active source, normally the FM Approved FISCO

barrier, is allowed to provide the necessary power for the Fieldbus system. All other

equipment connected to the bus cable has to be passive (not providing energy to the

system), except to a leakage current of 50μA for each connected device. Seperately

powered equipment needs a galvanic isolation to insure that the I.S. Fieldbus circuit

remains passive.

x

The cable used to interconnect the devices needs to comply with the following

parameters:

Loop resistance R’: 15 … 150 Ω/km; Inductance per unit length L’: 0,4 … 1 mH/km

Capacitance per unit length C’: 80 … 200 nF/km

(C’ = C’ line/line + 0,5 C’ line/screen if both line are floating)

(C’ = C’ line/line + C’ line/screen if the screen is connected to one line)

Length of spur cable: max. 30 m

Length of trunk cable: max. 1 km

Length of splice : max. 1 m

WARNING

- Substitution of components may impair Intrinsic Safety

- To prevent ignition of flammable or combustible atmospheres, disconnect power

before servicing or read, understand and adhere to the manufacturer’s live

maintenance procedures.

Application Doc. No.: IKE025-A10 P.6 to P.7