11-2. Wiring diagrams
11-2-1. Example of Non-Explosionproof System

15

PH8EFP
PH8ERP
PH8EHP

"2

WTB10

Terminal Box

PH202G

pH/ORP Transmitter

16
13
12

11
14

PH201G

Distributor

SDBT

Distributor

15
16
13

+

-

A(+)
B(-)

Output Signal
(1 to 5V DC)

Output Signal
(1 to 5V DC)

1(+)
2(-)

(+)A

(-)B

C

e

WASH

f

(CMN)D

G

12

11
14

c

FAIL

d

a

Hold

b

F2.3E.eps

(100Ω or less)

11-2-2. Example of Intrinsically Safe Explosionproof System

PH8EFP
PH8ERP
PH8EHP

PH202S

pH/ORP Transmitter

Safety Barrier

Distributor

15
16
13

+

-

Output

G

12

11
14

*1

pH/ORP
Sensor

*1: Use a 2-conductor shielded cable with an outside diameter of 6 to 12 mm. Shield must be connected to internal terminal G of transmitter

and left unconnected at the other side.

*2: Transmitter must be grounded using external terminal: for general purpose version ground resistance of PH202G should not exceed 100V

(Japanese Class D grounding) .

Ground to earth

F2.4E.eps

Use an appropriate DC power supply (such as from the PH201G distributor) for the PH202

transmitter. Under no circumstances should you connect AC power such as 100V AC or similar AC

power supply line. To measure pH or ORP in hazardous locations, use the PH202S or PH202SJ with

intrinsic safety barriers.

Grounding:

Be sure to ground the transmitter by using the ground terminal on its case.

Connect the G terminal inside the transmitter, to the shield wire of two-core shield cable which is

conneced between the distributor and transmitter.

For the PH202G transmitter (this does not apply to the PH202S) if you cannot ground the G

terminal on the transmitter case then connect this G terminal to the shield of the two-wire cable

connecting the transmitter and distributor, and ground it at the distributor end.

11-3. Sensor wiring

Refer to Figure 11-1 and 11-2, which includes drawings that outline sensor wiring.

The PH202 can be used with a wide range of commercially available sensor types, both from

Yokogawa and other manufacturers. The sensor systems from Yokogawa fall into two categories; the

ones that use a fixed cable and the ones with separate cables.

To connect sensors with fixed cables, simply match the terminal numbers in the instrument with the

identification numbers in the instrument on the cable ends.

The recommended procedure is to color-code each end of the cables to match the sensors with the

color strips provided with each cable. This provides a quick way to identify the ends of the cables

belonging to a particular sensor when they are installed.

(The procedure for fixing the identification labels is described in detail in the instruction sheet

provided with the cable.)

WARNING

WARNING