6-2. sensor cable connection with special grommet – Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 43

Advertising
background image

IM 12B07D02-01E

3-10 Installation and wiring

Figure 3-10b. Sensor wiring

3-6-2. Sensor cable connection with special grommet

In order to seal multiple sensor cables into EXA, a special grommet is provided that is designed to

accommodate one, two or three sensor cables (5 mm dia.) plus a liquid earth cable (2.5 mm dia.). In the

pack with the grommet are blanking pieces to close any unused holes. When correctly assembled, the

grommet maintains the IP65, NEMA 4X rating of the EXA PH202 housing.

Refer to figure 3-5 to assemble the grommet connections:

1. First remove the nut and standard rubber seal from the selected gland

2. Discard the seal. This will be replaced later by the special grommet

3. Thread the cables through the nut and the gland

4. Connect the cables to their designated terminals

5. Arrange the cables to avoid tangles and insert the grommet between the gland and the nut

6.

The grommet is split to permit the cables to be mounted after connection. (This also ensures even

length adjustment.)

7. Ensure that any unused holes are filled with the blanking pieces

8. Tighten the nut to form a firm seal. (Hand-tight is sufficient.)

NOTE:

The special gland is intended to be used to seal the multiple cables from the Yokogawa flow

fittings such as FF20 and FP20. The designated cables are WU20 sensor cables, which are

approximately 5 mm (0.2") in diameter, and 82895002 liquid earth cables, which are approximately

2.5 mm (0.1") in diameter.

For sensor systems using a single cable, like the FU20 (FU25) and the PR20, PD20, PF20 and PS20,

the standard gland will accommodate the cable adequately. Single cables between approximately 6 mm

and 12 mm (0.28" and 0.47") can be sealed properly with these glands.

*

*

*

*

*

Separate electrodes
ORP/REF electrode

Red

Yellow

Cable markers

11 Temperature
12 Temperature
13 Reference
14 Solution ground
15 Metal (measure)
16 Shield
17 Shield

Red
Black

Red

Blue
Blue

11 Temperature
12 Temperature
13 Reference
14 Solution ground(ORP)
15 pH
16 Shield
17

ORP/rH wiring

Note: ORP and or rH setting in service

code 02.

11 Temperature
12 Temperature
13 Reference
14 Solution ground
15 Metal (measure)
16 Shield
17 Shield

Blue

Blue

Red
White

Cable markers

Combined ORP/pH electrode

Notes
1. A temperature sensor maybe

connected to 11 & 12, for
temperature indication.

2. See section 5-3-1 for

impedance checking of glass
reference sensors.