4-4. setting of errors (fail contact), 4-4. setting of errors (fail contact) -9 – Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 106

Advertising
background image

IM 12B07D02-01E

Appendix 11-9

11-4-4. Setting of errors (FAIL contact)

Code 53 *Err.4.1-*Err .16 (9 items)

These set how errors are notified to the user. Either of two modes can be set. "Hard

Fail" mode lights the "FAIL" indicator on the LCD panel. When "enable status output

contact" is set for the PH201G distributor (Style B), the FAIL contact remains closed and

the "wash" contact stops operating. If Burnout functions are enabled by Code 32, then

the output becomes Burnout Upscale or Burnout Downscale.

"Soft Fail" mode blinks the "FAIL" indicator on the LCD panel. When "enable status

output contact" is set for the PH201G distributor (Style B), the FAIL contact pulses and

the "wash" contact and maintenance contact continue to operate normally. In "Soft Fail"

mode, notification that "periodic maintenance is due" still functions (service code 55) so

– for this reason – it is better to set "Soft Fail" mode than "Hard Fail" mode.

T5.3.5E.eps

Code Display Function

Function detail

X Y Z Ddefault values

User interface

53

*Err.4.1 Error setting

Impedance low (input 1) Soft fail

0

Impedance low (input 1) Hard fail

1

1

Hard

*Err.5.1

Impedance higt (input 1) Soft fail

0

Impedance high (input 1) Hard fail

1

1

Hard

*Err.4.2

Impedance low (input 2) Soft fail

0

Impedance low (input 2) Hard fail

1

1

Hard

*Err.5.2

Impedance higt (input 2) Soft fail

0

Impedance high (input 2) Hard fail

1

1

Hard

*Err.07

Temperrature too high Soft fail

0

Temperrature too high Hard fail

1

1

Hard

*Err.08

Temperrature too low Soft fail

0

Temperrature too low Hard fail

1

1

Hard

*Err.09

pH out of range Soft fail

0

pH out of range Hard fail

1

1

Hard

*Err.11

Wash recovery check

Soft fail

0

0

Soft

Wash recovery check

Hard fail

1

*Err.16

Call for maintenance

Soft fail

0

0

Soft

Call for maintenance

Hard fail

1

Advertising