10-4. Set up for other functions
• Current

outputs

Transmission signals for the measured parameters and FAIL signals can be set up in service codes

31, 32 and 35.

• Diagnostic

Checks

Impedance checks, response time and stability checks are all included in the PH202. In order to get the

best performance from each of these features, the transmitter should be fine tuned according to experi-

ence in the installation, and for the particular sensors selected. Service codes 3, 4, 5 & 20 all contribute to

the diagnostics. Please note that the default settings provide an excellent starting point and provide most

valuable information about the performance of the electrode system.

• Communications

The proprietary HART (FSK) communication link allows remote configuration and data retrieval

through the PC202 communication package. This is an excellent tool for the maintenance engineer,

quality engineer or plant manager. Service codes 60-62 are used to set up the communications.

• Logbook

In combination with the communications link, a “logbook” is available to keep an electronic record of

events such as error messages, calibrations and programmed data changes. By reference to this

log, users can easily evaluate diagnostic information to determine predictive maintenance schedules.

For example, by monitoring the deterioration in the slope of the pH sensor, it can be changed before

a failure (or process shutdown) occurs.