Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 67

Advertising
background image

IM12B07D02-01E

5-16 Parameter setting

Code Display

Function

Function detail

X Y

Z Default values

Calibration functions
20 *∆t.SEC

Stability check time

5 sec.

*∆PH

Stability check pH

0.02 pH

21

*AS.LOW

As Pot low limit

-120 mV

(As Pot) *AS.HI

As Pot high limit

120 mV

21

*ZP.LOW

Zero Point low limit

5.00 pH

(Zero) *ZP.HI

Zero Point high limit

9.00 pH

22

*SL.LOW

Slope low limit

70 %

*SL.HI

Slope high limit

110 %

23

*ITP

Set ITP

Preset calibration data from manufacturer

7.00 pH

(pH)

*SLOPE

Set slope

or from laboratory determinations.

100 %

*ASP.1D

Set As Pot

For the main parameter

0.0 mV

Press YES to confirm 0.1 mV resolution,
then set value with >, ^, ENT keys.
Press NO to change to *ASP.

*ASP

Set As Pot

For the main parameter
Press YES to confirm 1 mV resolution,
then set value with >, ^, ENT keys.

*ASPmV

Set As Pot ORP

For parameter 2 (when activated in

service code 02)

23

*ASP.1D

Set As Pot (ORP)

For the main parameter

(ORP)

Press YES to confirm 0.1 mV resolution,
then set value with >, ^, ENT keys.
Press NO to change to *ASP.

*ASP

Set As Pot

For the main parameter
Press YES to confirm 1 mV resolution,
then set value with >, ^, ENT keys.

24

*BUF.ID

Buffer table 4

Buffer tables to NIST (formerly NBS)

25

*BUF.ID

Buffer table 7

(see section 10 for table details)

26

*BUF.ID

Buffer table 9

User adjustable for special requirements

27

*ZERO.P

Enable zero point in Disable zero point (enable As Pot)

0

0 Disabled

pH units

Enable zero point (disable As Pot)

1

28-29

Not

used

Fig. 5-1.

pH

mV

ITP

As pot

0 mV

0 °C
10 °C
25 °C

mV

500

0

- 200

14

pH

ITP

Zero Point

0

7

Select YES/NO

Select YES/NO