6-1. connection cable – Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 42

Advertising
background image

IM 12B07D02-01E

Installation and wiring 3-9

3-6-1. Connection cable

There are two types of connection cable, one for single sensors and one for combined sensors. The

former is a coaxial cable and has only two connections.

• Red

to measuring element

• Blue

to screen (shield)

The latter is a triaxial cable with three connections, (it has an extra white wire termination) these wires

are connected:

• Red

to measuring element

• Blue

to reference

• White to screen (shield)

To connect the other sensor systems, follow the general pattern of the terminal connections as listed

below:

11 & 12

Temperature compensation resistor input

13

Input no. 2 (normally the reference element)

17

Screen (shield) for input no. 2

14

Liquid earth (solution ground) connection

15

Input no. 1 (normally the measuring element)

16

Screen (shield) for input no. 1

*

*

*

*

*

*

*

pH transmitter

Green

Yellow

Red

Cable markers

11 Temperature
12 Temperature
13 Reference
14 Solution ground
15 Glass (measure)
16 Shield
17 Shield

Red
Blue

Red

Black
Red
Blue
Blue

11 Temperature
12 Temperature
13 Reference
14 Solution ground
15 Glass (measure)
16 Shield
17 Shield

11 Temperature
12 Temperature
13 Reference
14 Solution ground
15 Glass (measure)
16 Shield
17

11 Temperature
12 Temperature
13 Reference
14 Solution ground
15 Glass (measure)
16 Shield
17 Shield

Combined pH/ref sensors

Cable markers

Red
Blue
Blue

Black
Red
White

Green

Blue

FU20/FU25 4-in one sensor

Note: Connect cables to similarly marked
terminals: 11 to 11, 12 to 12, etc.

Retractable sensor assembly PR20
(Also PD20, PF20 & PS20)

NOTE:

Connect cables to similarly marked terminals:

11 to 11, 12 to 12, etc.

Figure 3-10a. Sensor wiring