Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 27

Advertising
background image

IM 12B07D02-01E

2-14 Specification

2-11. Control Drawing of PH202S FF/PB Specification (FM Intrinsically safe Entity).

Unclassified Location

Classified Location

Division 1

FM Approved

barrier

Voc (Vt) d 24 V

Ioc (It) d 250 mA

Poc (Pt) d 1.2 W

Ca t 220pF+ Ccable

La t 0 H + Lcable

Sensor

Connections

Max. cablelength: 60 mtr.

Cable dia. : 3…12 mm.

FM Class I, DIV. 1, Group ABCD

T4 for ambient temp. d 55 qC

-

+

I.S.

certified

Terminator

I.S.

certified

Terminator

PH202S-F

or PH202S-P

Transmitter

Transmitter

-

+

Sensor

Connections

-

+

-

+

x

Sensor(s) are of a passive type to be regarded as 'simple apparatus', devices which neither

store nor generate voltages over 1.5 V, currents over 0.1 A, power over 25 mW or energy over

20 PJ, or are FM Approvals entity approved and meet connection requirements.

x

Electrical data of the PH202S-F & PH202S-P:

-

Supply

circuit:

Maximum input voltage Vmax = 24 V

Maximum input current Imax = 250 mA

Maximum input power Pi=1.2 W

Effective internal capacitance Ci = 220 pF;

Effective internal inductance Li = 0 PH.

-

Sensor input circuit:

Maximum output voltage Vt = 14.4 V;

Maximum output current It = 32.3 mA

Maximum allowed external capacitance Ca = 600 nF

Maximum allowed external inductance La = 34 mH

x

Any FM Approved barrier may be used that meets the following requirements:

Voc or Vt d 24 V

Ioc or It d 250 mA

Poc or Pt d 1.2 W

Ca t 220 pF + Ccable; La t 0

μH + Lcable

When installing this equipment, follow the manufacturer’s installation drawing.

Installation should be in accordance with ANSI/ISA RP 12.06.01 “Installation of Intrinsically

Safe Systems for Hazardous (Classified) Locations” and the National Electrical Code

(ANSI/NFPA 70).

Associated apparatus connected to the barrier must not use or generate more than

250 Vrms or Vdc.

x

Resistance between Intrinsically Safe Ground and earth ground must be less than 1.0 Ohm.

x

In case of using cable glands in Outdoor location, they shall be UV rated or made of metal.