5-2. manual calibration, 5-3. calibration with sample, 5-4. data entry – Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 108: 5-2. manual calibration -11, 5-3. calibration with sample -11, 5-4. data entry -11, Warning

Advertising
background image

IM 12B07D02-01E

Appendix 11-11

(6) After the display reading has stabilized, read and note it. Measure the temperature of the test

solution, and check the ORP of the test solution at that temperature using Fig. 11-3.

If the ORP reading is within the tolerance shown in Fig. 11-3 then the sensor is normal.

Note: If the measurement value is only slightly out of tolerance, you should check whether the test

solution was fresh and whether it was mixed in the correct amounts (see mixing instructions

above).

0

10

20

30

40

50

0

100

200

300

400

500

600

(mV)

Quinhydrone solution

ORP

Temperature C

Tolerance

Tolerance

Ferrous & Ferric solution

F6.1E.eps

Fig. 11-3 ORP of Test Solution

(7) Replace the sensor in its holder.

Be careful to make sulfuric acid solution.

11-5-2. Manual Calibration

The procedure for calibration is described in 11-5-5 (1).

Calibration is performed by using a solution of known ORP value, and adjusting the PH202 to read

the same value. Alternatively you may calibrate the PH202 to read the same as another 'standard'

instrument in a given solution. The electrode check described in 11-5-1 above is performed if the

measured value appears to be well out of tolerance. Calibration of the ORP value is performed by

zero-shift one-point calibration.

11-5-3. Calibration with Sample

The procedure is described in 11-5-5 (2).

Calibration with a sample enables measuring a representative sample (e.g. by laboratory analysis)

while the ORP sensor is on line and measuring the same sample solution, saving the measured data

in the memory of the PH202, and then adjusting the PH202 to read the same value (as obtained

e.g. by laboratory analysis). So normal on-line control or pH measurement is not affected. The ORP

measurement is calibrated by zero-shift one-point calibration.

11-5-4. Data Entry

If the above methods of calibration present problems, another alternative is the direct entry of

calibration data using the Service Code Menu (refer to Chapter 5). For example, the sensor may

be calibrated in a test laboratory and then transported to its location in the field and the sensor

parameters entered into the associated PH202.

Service Code 23 allows you to access and enter ITP, Asymmetry Potential (or Zero Point) and Slope

parameters.

WARNING