Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 127

Advertising
background image

IM 12B07D02-01E

Appendix 12-15

3/3

QIS 12B07D02-71E

3.6 Profibus Communication Functional Check

Check for normal function using Profibus equipment specified by Yokogawa.

In the tests of Item 3.3 to 3.5, check the communication function using Profibus communication.

PH202

+

G

11

12

14

17

13

LOW IMP

-

+

Decade

Resistance Box

Standard

Voltage Source

Shorting Bar

Profibus Equipment Specified by YOKOGAWA.

C

R

C

R

R=50

Ω± Ω

C=2μF±0.4μF

SUPPLY

TEM

P

LE

HIGH IMP

15

16

INPUT2

INPUT1

Ground

+

DA/PA Coupler

DC

Source

24VDC

Figure 1 Testing Circuit and Test Equipment