3. wiring of sensors, 3-1. general precautions – Yokogawa EXA PH202 2-Wire pH/ORP Analyzer User Manual

Page 37

Advertising
background image

IM 12B07D02-01E

3-4 Installation and wiring

Figure 3-6. System configuration

3-3. Wiring of sensors

3-3-1. General precautions

Generally, transmission of signals from pH sensors is at a very low voltage and high impedance level.

Thus a lot of care must be taken to avoid interference. Before connecting sensor cables to the transmit-

ter make sure that next conditions are met:

– the sensor cables are not mounted in tracks together with high voltage and or power switching

cables

– only standard coaxial electrode cables or extension cable are used

– the transmitter is mounted within the distance of the sensor cables (max. 10 m)

– the setup is kept flexible for easy insertion and retraction of the sensors in the fitting.

3-3-2. Additional precautions for installations in hazardous areas

Make sure that the total of capacitance and inductances connected to the input terminals of the EXA

PH202S do not exceed the limits given in the certificate.

This sets a limit to the cable and extensions used.

– The intrinsic safe version of the PH202S instrument can be mounted in Zone 0 or 1 (PH202S-B, -N,

-D).

– The sensors can be installed in Zone 0 or Zone 1 if a safety barrier according to the limits given in

the system certificate is used.

– Ensure that the total of capacitance and inductances connected to the terminals of the EXA

PH202S do not exceed the limits given in the certificate of the safety barrier or distributor.

– The cable used should preferably have a BLUE colour or marking on the outside.

– Installation for (sensors in Zone 0 or 1):

Generally, the distributor with input/output isolation has no external earth connection. If there is an

earth connection on the distributor and the external connection of the transmitter is connected to

“protective” earth, the shield of the 2-wire cable may NOT be connected to “protective” earth at the

distributor too.

1 2

1 8 0

1 0 0

0

NO

MODE

YES

>

ENT

>

F AIL

HOLD

YES NO

ENT

MEASURE

MAN.CAL

DISPLAY

HOLD

YOKOGAWA

MODE

TEMP

AUT.CAL

TEMP.MAN.

pH

Sensors

Recorder

Computer

Hand Held
Communicator

Output/supply

Input

2 , 5 or 10 m

Safety Barrier
PH202S only

CURRENT OUTPUT

ref pH/ORP emp.

Distributor

Advertising