2 peg test – Spectra Precision Survey Pro v4.6 Ranger User Manual User Manual

Page 207

Advertising
background image

Leveling

197

2 Peg Test

The 2 Peg Test is used to check the collimation error of the
instrument. The test will compute the error, which can then be used

to adjust the horizontal crosshair of the instrument.

1. Position two rods 50 to 90 meters apart (165 to 300 feet).

2. Pace off the distance between the rods and set up the level

midway between them. (The placement of the rod over Point

B can be adjusted after shooting Point A.)

3. Carefully level the instrument. You should be able to rotate

the instrument 180° around its vertical axis without the
bubble moving away from the center.

4. From the Main Menu, select Leveling > 2 Peg Test .

5. Tap Take Shot. The Leveling Shots

screen will open where you will be

prompted to take each set of shots to
the rod over Point A. Once each shot is

completed for the number of sets

entered in the Level Settings screen,
you will return to the Peg Test screen.

6. Turn the instrument to the rod over

Point B.

7. You can optionally tap Check H. Dist to

verify the instrument is centered

between the two rods. The distance to Point B will be

measured and compared to the horizontal distances

previously measured to Point A and a come/go distance will be
provided so the rod at Point B can be moved to equal the

horizontal distance to Point A prior to taking the shots that

will be used to compute the error.

8. Tap Take Shot. The Leveling Shots screen will open again

and prompt you to take the required shots to the rod over
Point B. Once completed, the Peg Test screen will open with

a new graphic showing the next setup.

Advertising