Section 2: System Components

RuggedCable® System
Cable includes conductors for power and communication

signals, a weight-bearing structure, and a Kellems® grip

to anchor the Aqua TROLL Instrument securely. Cable is

available in standard and custom lengths.

Uphole and downhole ends are identical female

twist-lock connectors that connect to the Aqua TROLL

Instrument, cable connect TROLL® Com devices,

desiccants, and other accessories.

Vented cable must be used with vented pressure/level

sensors on the Aqua TROLL 200 to achieve gauged

measuresments. The cable vent tube ensures that

atmospheric pressure is the reference pressure applied to

the sensor diaphragm.
Vented cable ships with a small desiccant cap that

should be replaced with a larger volume desiccant

before you deploy the instrument in a humid

environment.
Non-vented cable may be used with non-vented

pressure/level sensors on the Aqua TROLL 100 or 200

Instruments to achieve absolute measurements. Vented

cable can also be used.