Maintenance and recalibration – In-Situ Aqua TROLL 100 and 200 Operators Manual User Manual

Page 57

Advertising
background image

Page 57

Aqua

TROLL Operator’s Manual

Maintenance and Recalibration
The need to clean the sensor is determined by the site

conditions, and the accuracy you want to obtain from the

instrument. For example, when used in relatively clean water of

low biological productivity, a conductivity sensor could maintain

its accuracy specifications for several months. On the other hand,

in water with a high potential for biological or mineral fouling,

the sensor may need to be cleaned and recalibrated more

often. Visual inspection and measurement results are the best

indicators that a device needs service recalibration.

It is the user’s responsibility to develop and follow a maintenance

program appropriate to the site. Most standard operating

procedures (SOPs) instruct that the instrument be checked in a

known standard at the end of a deployment. Calibration should

always be checked after cleaning, and should be recalibrated

when necessary.

For specific cleaning procedures, refer to the Care and

Maintenance recommendations in

Section 9.

Factory recalibration of the Aqua TROLL 200 Instrument should

be performed every 12-18 months. Contact In-Situ Inc. as

described on

page 10

.

Section 5: Conductivity