Desiccant
Vented cable must be installed with

outboard desiccant to protect the

cable vent tube and Aqua TROLL

electronics from condensation in

high-humidity environments.

The desiccant may be removed from

the vent tube, if needed, to trim the

conductor wires. Pull the vent tube

extender off the cable vent tube

to remove, replace desiccant after

trimming and connecting the wires.

Wiring
Refer to diagrams on the following pages. Trim back and insulate

unused wires. The shield should be wired to a chassis ground or

earth ground.

and secured to

cable with strap

Outboard desiccant

is attached to cable

vent tube

RuggedCable

Signal Color Pin

Gnd/Return BLACK

6

Ext Power RED

5

4-20 mA

BROWN

4

RS485(–) GREEN 3
RS485(+) BLUE 2
SDI-12 WHITE 1

m

2

m

1

f

5

m

3

f

6

f

4

Section 8: Analog, SDI-12, Modbus