Section 9: Care & Maintenance

Storage
Store the Aqua TROLL Instrument clean and dry. Place the

protective red dust cap on the cable end, or store with cable

attached to protect the connector pins and O-ring.

Store the instrument where it will not fall onto a hard surface.

Protect the instrument from temperature extremes. Store within

a temperature range of -40° C to 65° C (-40° F to 149° F).

General Maintenance
Twist-Lock Connectors
Keep the pins on all connectors free of dirt and moisture by

using the soft protective dust cap when cable is not attached to

an instrument.
Cable Vent Tube (Vented Cable)
Vented cable ensures that atmospheric pressure is the reference

pressure to the vented pressure sensor diaphragm. The vent

tube should not be blocked, kinked, or otherwise obstructed.

Such obstructions will cause barometric pressure to appear in

measurements, and errors will be introduced due to thermal

expansion and contraction of air within the vent tube and probe.

The recommended minimum bend

radius is 13.5 mm (0.54 in), which is

twice the cable diameter.

The minimum
bend radius for
vented cable is

13.5 mm (0.54 in).

Do not sub merge
the cable
connector; do not

immerse in any fluid.

TIP: Be sure to
replace the desiccant
before it appears

pink. Expired desiccant can
allow moisture build up in the
vent tube, causing a blockage
resulting in inaccurate data.