Section 5: Conductivity

Available Parameters
The Aqua TROLL Instrument can also calculate and record

derived parameters as listed below.

For more information on measurement methodology, including

the equations used to derive the calculated parameters, see the

technical note “Aqua TROLL 200 Measurement Methodology” on

the In-Situ software/resource CD, or in the Downloads section of

the In-Situ web site at www.In-Situ.com.
Actual Conductivity
Units: micro siemens per centimeter (µS/cm), default

milli siemens per centimeter (mS/cm)

Specific Conductivity
A means of expressing what the actual conductivity of a solution

would be at a standard reference temperature (25° C). Calculated

from actual conductivity and temperature. The factory default

coefficients calculate specific conductivity per Standard Methods

2510B.

Units: micro siemens per centimeter (µS/cm), default

milli siemens per centimeter (mS/cm)

Salinity
Calculated from actual conductivity and temperature. Units:

Practical Salinity Units (PSU)
Total Dissolved Solids (TDS)
Calculated from specific conductivity using a conversion factor

(Default: 0.65).

The TDS conversion factor may be edited in Win-Situ as follows:

Select the sensor on the Sensors tab, click Configure

, in