9 Care & Maintenance

Operating Considerations
The Aqua TROLL Instrument has been designed to withstand

harsh field conditions. However, as with any electronic

instrument, it can be permanently damaged if used outside its

operating specifications.
Temperature
The Aqua TROLL temperature sensor operates within a

temperature range of -5° to 50° C (23° to 122° F) in non-freezing

fluid.
Pressure Range
The Aqua TROLL 200 Instrument can withstand pressures of up

to two times (2X) the rated range of the pressure sensor without

damage, although it may not read correctly at such pressure. If

the pressure range is exceeded by 3X, the pressure sensor will be

destroyed.
Factory Calibration
Accuracy can be adversely affected by improper care and

handling, lightning strikes and similar surges, exceeding

operating temperature and pressure limits, physical damage or

abuse. Factory calibration every 12-18 months is recommended.

Contact In-Situ Customer Service for information on the factory

maintenance and calibration plan.

Do not deploy
pressure
transducers such

that ice may form on or near
the sensor or cable connections.
Ice formation is a powerful
expansive force and may
over-pressurize the sensor or
otherwise cause damage. Any
damage associated with ice
formation is not covered by the
warranty.