In-Situ Aqua TROLL 100 and 200 Operators Manual User Manual

Page 50

Advertising
background image

Page 50

Aqua

TROLL Operator’s Manual

A Eliminate water and rinse
1. Remove the nose cone (Aqua TROLL 200 only). Water trapped

here can dilute the calibration solution. Air bubbles may also

come from this area.

2. If the device is wet from previous use, dry the body and shake

to clear any liquid inside the conductivity sensor.

3. Before opening the solution bottle, invert it a few times to

redistribute any water condensation.

4. Remove the Cal Cup cap and fill the cup to the “Rinse” line

with calibration solution.

5. Insert the Aqua TROLL through the grommet in the cap.
6. Attach the cap to the Cal Cup. The Aqua TROLL

should rest on or near the bottom.

7. Shake vigorously to rinse the sensing cell.
8. Loosen the cap, remove the Aqua TROLL. There is no

need to pull it out of the cap. Discard the solution.

9. For best results, rinse again using the same

procedure.

B Equalize temperature
1. Fill the Cal Cup to the “Fill” line with calibration

solution.

2. Install the Aqua TROLL and tighten the cap.
3. Invert multiple times for at least 30 seconds—

longer if the probe and solution are at different

temperatures.

Section 5: Conductivity

When the nose
cone is removed,
the pressure sensor

diaphragm is exposed. Do not
touch the the pressure sensor
membrane.

Cal Cup

Grommet

Fill

Rinse

Cap

TIP: A second
rinse is important if
the Cal Cup and the

Aqua TROLL were not clean and
dry at the start of this process.