In-Situ Aqua TROLL 100 and 200 Operators Manual User Manual

Page 54

Advertising
background image

Page 54

Aqua

TROLL Operator’s Manual

If the cell constant is out of

range, the cause could be

an air bubble, incomplete

rinsing, sensor fouling, or

other factors. Repeat the

preparation and calibration

with fresh cal solution.

9. When you are satisfied with

the proposed cell constant,

click Commit to write the

calculated cell constant to

the sensor.

10. If the Aqua TROLL will be

deployed immediately,

remove it from the Cal Cup, rinse it, discard the calibration

solution, and reinstall the black nose cone. Otherwise,

store it wet in the Cal Cup (in cal solution or water) for later

deployment.

TIP: If the proposed
cell constant is
outside the range

0.98 – 1.02, check conditions
and repeat the calibration.

Section 5: Conductivity

Nominal Stability vs. Full Stability

To meet the criteria for a valid calibration point, the change in sensor
response is monitored over time. The software is looking for the
calibration solution temperature and the sensor readings to settle
over a specific time period. The criteria for Full Stability are designed
to meet the published specifications. The criteria for Nominal Stability
are designed to shorten the calibration time when an approximate
calibration is acceptable.

TIP: A change
from an earlier cell
constant of 0.02

suggests that the calibration
should be repeated.

CALIBRATION

(continued)