Section 10: Troubleshooting

Problem: Real-time readings are in the wrong units.
Probable Cause: Default units are being used.
Suggested Remedy: Click the Sensors tab, select the sensor,

click the Configure button

, and select the desired

units for each parameter in the Sensor Setup window. Click

OK

. Be sure to stop “polling” in the Home screen

before selecting units.

TROUBLESHOOTING DATA LOGS
Problem:
Cannot add a new log.
Probable Cause 1: Only one active log can reside in the device

at a time—an active log is a log that is Ready, Pending,

Running, or Suspended as shown in the Status column of the

Logging Tab.

Suggested Remedy: Stop or delete the log if possible.

Alternatively, configure the new log after the active log is

completed.

Probable Cause 2: The device has its maximum number of logs

already stored—although this is not a likely cause for the

Aqua TROLL 200, which has a capacity of 50 logs.

Suggested Remedy: Download, and then delete a log you are

finished using. This will make room for an additional log on

the device.

Problem: New log exceeds available memory (message from

software).

Probable Cause: The log as configured would exceed the device

memory.

Suggested Remedy: Edit the log and try these:

• Select a longer sampling interval.

• If available, select the Wrap data option (current data

will overwrite older data when the memory is full).