• Rinse the Aqua TROLL Instrument and the Cal Cup

thoroughly with fresh calibration solution before

calibrating.

Problem: The calculated cell constant (Kcell) is higher than the

recommended range of 0.98 – 1.02.

Probable Cause: The electrical resistance of the calibration

solution was higher than expected for the nominal value,

or an air bubble was present in the sensing cell during

calibration.

Suggested Remedy: Check the following and then repeat the

calibration:

• Visually inspect the sensor for the presence of fouling

(most likely an obstructing material or organism); clean

as described in Section 9

of this manual.

• Ensure the calibration solution is not contaminated,

mislabeled, or partially diluted.

• Rinse the Aqua TROLL Instrument and the Cal Cup

thoroughly with fresh calibration solution before

calibrating.

• “Roll and tap” the Aqua TROLL Instrument in the Cal

Cup to dispel air bubbles as described and illustrated in

Section 5.