Pump priming – Goulds Pumps IC - IOM User Manual

Page 48

Advertising
background image

Conditions for application of an external liquid

If...

Then...

The stuffing box pressure is above atmospheric

pressure and the pumped fluid is clean.

Normal gland leaks of 40 to 60 drops per minute is

usually sufficient to lubricate and cool the packing.

You do not need sealing liquid.

The stuffing box pressure is below atmospheric

pressure or the pumped fluid is not clean.

An outside source of clean compatible liquid is

required.

An outside source of clean compatible liquid is

required.

You must connect the piping to the lantern ring

connection with a 40 to 60 drops-per-minute leak

rate. The pressure must be 15 psi (1.01 kg/cm

2

)

above the stuffing box pressure.

Pump priming

Prime the pump with the suction supply above the pump

1. Slowly open the suction isolation valve.
2. Open the air vents on the suction and discharge piping until the pumped fluid flows out.
3. Close the air vents.

1

2

3

1.

Discharge isolation valve

2.

Check valve

3.

Suction isolation valve

Prime the pump with the suction supply below the pump

Use a foot valve and an outside source of liquid in order to prime the pump. The liquid can come from
one of these sources:

• A priming pump
• A pressurized discharge line
• Another outside supply

1. Close the discharge isolation valve.
2. Open the air vent valves in the casing.
3. Open the valve in the outside supply line until only liquid escapes from the vent valves.
4. Close the vent valves.
5. Close the outside supply line.

Commissioning, Startup, Operation, and Shutdown

46

Model IC, ICI, ICH, ICIH Installation, Operation, and Maintenance Manual

Advertising