Additional display – RADWAG MA 200.3Y Moisture Analyzer User Manual

Page 121

Advertising
background image

- 121 -

Enter the database of operators and edit desired operator’s record, go to <

Card number> option,

On entering editing mode in <

Card number> option, an editing window

<Card number> is opened with an on-screen keyboard,

Approximate the card to the transponder card scanner. The moisture

analyzer automatically displays number of detected card in the <Card

number> field,

Accept recognized number by pressing

key and returning to the main

window of weighing mode.

21.5. Additional display

The moisture analyzer enables communicating with an additional display using

the following ports:

RS 232 (COM1),

RS 232 (COM2),

Tcp.

Type of data displayed on an additional display is declared in variable <Project

(template)>.

Procedure – designing a variable

Enter group of parameters <

Peripheral devices>

Press “

Additional display /

Project” field, which opens an editing

window <Project> with an on-screen keyboard,

Use the on-screen keyboard to insert desired value of a standard, or

Select desired value of a standard using a list displayed on pressing <

> key

Accept by pressing

key.

Caution:
Default setting of parameter <

Standard>: a variable with value {140}.

Advertising