18.2.2. Use of menu <Settings>

After entering the moisture analyzer settings, the user can define all drying

parameters, and main data for printing completed measurements (data type,

time interval between printing the following measuring data). The data is

definable in one menu window.

Procedure:

1. Press grey coloured workspace

2. Moisture analyzer’s display indicates menu: Settings, Keys, Information,

Printouts, Profile,

3. Press <Settings> key,

4. The display indicates functions related to drying mode, such as:

Drying mode [means of heating a sample]

Finish mode [criteria on finishing drying process]

Result [moisture content measuring unit, active on drying process

start]

Printout interval [time intervals in which measurement data is printed]

Start mass control [enabled / disabled – meeting a requirement of

determined start mass of a sample, with declared tolerance of this mass

value (setting mass value and its tolerance is carried out in database of

products)]

Print measurements [printing indirect measurement result of moisture

content determining process, with time interval set in parameter

<Measurement time >]

Means of setting each parameter is the same as in case of the previous

menu option. The only difference refers to possibility of declaring parameter

of controlling start mass for a sample.

This parameter determines use of a sample with specific mass to test its

moisture content.

CONTROL OF START MASS enables setting one of three options:

No – option disabled, mass of a sample prepared for drying is not controlled

by the software

Optional – option enabled, the display shows a bar graph demonstrating

graphic visualization of sample mass in relation to mass value saved in the

database of products. It is a sample mass and its tolerance limits set in the

software.

Before applying this option, add a product in the database, including data on

sample mass and its checkweighing limits.

On enabling this option, the moisture analyzer’s software displays

commands on placing a sample with determined mass on the drying pan.

However, if a user decides otherwise, then a moisture analyzer starts drying

process even if sample mass exceeds determined threshold limits for a

tested sample mass.