31. MOISTURE ANALYZER MAINTENANCE ACTIVITIES

This section of the user manual describes procedures of maintaining a moisture

analyzer MA 3Y series in good condition, and means of replacing faulty

components (fuses).

31.1. Cleaning moisture analyzer components

In order to ensure required measuring accuracy, the moisture analyzer has to be

used and stored in clean conditions.

While cleaning the moisture analyzer, precautions provided in this section of the

user manual must be obeyed.

REMEMBER, before initiating any maintenance or cleaning

activities of moisture analyzer’s inner parts, switch off the moisture

analyzer by pressing ON/OFF key on terminal’s overlay, and make

sure the power cord of the moisture analyzer is unplugged from

mains!

Open moisture analyzer’s lid and take

out all components of the drying

chamber:

disposable pan (5),

pan bracket (4),

pan handle (3),

drying chamber’s cover (2),

insert of the drying chamber’s basis

(1).

Clean the moisture analyzer using soft

fabrics (diaper, chamois, etc.) and

non-aggressive cleaning agents.

Do not use any abrasive agents or

solvents to clean the moisture

analyzer, as it may cause damage to

the device and its components.

After cleaning, thoroughly dry each of moisture analyzer’s components.

Do not allow any contamination or liquid to fall into the moisture analyzer.

Assembly cleaned and dried components in accordance with above scheme.