Before starting the temperature adjustment process, close drying chamber lid

and follow below process descriptions:

Enter submenu group

<Adjustment>.

Start <Temperature

adjustment> application by

pressing field with its name.

The moisture analyzer displays

a message box on procedure

start, and data on the process

(step number, current

temperature value, heating

time, burn-in (heating) time).

After 8 minutes, the system

opens a window with an on-

screen numeric keyboard.

Using the on-screen keyboard

insert the current temperature

value of the moisture analyzer,

that is indicated on the

external control thermometer.

Press <Accept> key to proceed

to the following adjustment

step.

The moisture analyzer switches

on the halogen heater and for

another 8 minutes it heats the

drying chamber and

simultaneously displays current

temperature value and other

parameters.