2.2. Safety information

CAUTION!

The use of moisture analyzer MA 3Y regardless of both, safety

information and user manual guidelines may cause health damage

and even death

WARNING:

Nominal voltage for a moisture analyzer is 230 VAC. It means

that safety usage precautions for low voltage devices must be

abide by while operating the device.

Three-core power supply cable with grounding pin comes

standard with a moisture analyzer.

If necessary, an extension cord can be used as long as it meets

the applicable standards and has a protective ground conductor.

Intended disconnection of the grounding cable is forbidden.

CAUTION:

Drying chamber cannot be opened during operation (drying

process). It is due to the fact that a halogen lamp and its glass

shield may reach the temperature of up to 400 C.

When setting up the moisture analyzer leave enough space to

prevent heat from building up and to keep your analyzer from

overheating. Leave about 20 cm around the instrument and

about 1 m above.

Air vents that are located in the housing cannot be covered,

sealed or blocked in any other way.

Do not put any flammable substances on, under or near the

moisture analyzer.

Be particularly careful when removing the sample from the

drying chamber: the sample itself, the drying chamber, shields

and the pan can still be extremely hot.

In case of any maintenance work (cleaning the inside of the

drying chamber), the moisture analyzer must be switched off.

Wait until all the components have cooled down. Do not perform

any modification to the heating module.

Same types of samples require taking particular safety

precautions.

They can pose a danger for people and objects. It is always the user

who is liable for possible damages caused by the use of an

inappropriate sample.