RADWAG MA 200.3Y Moisture Analyzer User Manual

Page 142

Advertising
background image

- 142 -

Format: BP_TIME <CR><LF>

Accessible responses:

BP_OK <CR><LF>

– command completed, BEEP signal activated

BP_E” <CR><LF>

– incorrect format or lack of the parameter

BP_I <CR><LF>

– command understood but not accessible at this

moment

TIME – parameter, decimal value, determined time limit for the signal, given

in [ms]. Recommended range <50

÷

5000>.

If value greater than the valid one is given, the BEEP sound is activated for

the maximal available time limit.

Example

Command:

BP_350<CR><LF> – turn BEEP signal for 350 ms

Response:

BP_OK<CR><LF> – BEEP on

Caution! – BEEP signal activated by BP command gets aborted if it is activated

by means of other devices: keyboard, touch panel, proximity sensors.

OMI - list available working modes

The command lists available working modes for a given device

Format: OMI <CR><LF>

Accessible responses:

OMI <CR><LF>

n_”Mode name”

<CR><LF>

:

n_”Mode name”

<CR><LF>

OK <CR><LF>

– command completed, available working modes are

given

OMI_I <CR><LF>

– command understood but not accessible at this

moment

Mode name – parameter, working mode name, given in between inverted

commas. Form of the name is the same like on the balance display, it is given

in a currently operating language.

n – parameter, decimal value specifying working mode number.

n → 1 – Weighing

17 – Pippette calibration

Caution: working modes numbering is assigned to working modes names and

it is the same for each balance types.

Some balances in response to OMI may give working modes numbers without

names.

Example1

Command:

OMI <CR><LF>

– give available working modes