When assembling the pan

supporter (bracket) pay special

attention to the correct

positioning of its mandrel. The

mandrel features a cut for its

unique positioning against the

pan handle, thus preventing their

contact and friction.

Setting pan’s bracket:
-

On assembling the pan

bracket onto the mandrel,

turn the bracket slightly, so

that the cut on the mandrel

is located in their unique and

correct position,

-

When turning the mandrel use

most gentle movements not

to damage moisture

analyzer’s measuring system.

4.3. Level setting

The moisture analyzer MA 3Y series features an electronic level control

system (and thus no external level). After software startup enter the leveling

application and follow moisture analyzer prompts to carry out the leveling

procedure.

Means of moisture analyzer leveling is described further in this user manual

(see point 11.3).

4.4. Standard delivery components

A moisture analyzer MA 3Y series.

Insert of drying chamber basis.

Drying chamber shield.

Drying pan handle.

Drying pan bracket.

Disposable pan.

Power cord.

Protective cover

Cable P0108 for connecting moisture analyzer and computer

User manual (CD).