Compare result obtained on a moisture analyzer with the one of

traditional method

When drying a sample with high

moisture content, it is possible to

shorten measurement time by

selecting “quick” drying mode. In

such case, the majority of moisture

content

is

released

drying

temperature higher than set. Only

after some time, the temperature is

lowered to the set value, and

maintained until completing drying process.


17.2.4. Selecting drying mode

The software of moisture analyzer MA 3Y series enables selecting one of four

drying modes:

Standard

Quick

Mild

Step

STANDARD mode
Standard mode is the most frequently used of

all drying modes. It enables accurate

determining of moisture content is a dried sample.

MILD mode
Mild mode is used in case of drying substances

that are sensitive to rapid heat emitted by

filaments operating at full power in the initial

stage of drying process. This mode prevents from

decomposing of substances sensitive to heat by

mild temperature increase in set amount of time

(time interval has to be selected by tests). Mild mode is recommended to

drying samples of leather structure.

QUICK mode
Quick mode is recommended to drying samples

that moisture content varies from 5%, and 15%.

In the initial stage of Quick mode, heating takes