RADWAG MA 200.3Y Moisture Analyzer User Manual
Page 29

- 29 -
Access level
Software of moisture analyzer MA 3Y series provides four access levels:
administrator, advanced operator, operator and none (anonymous logging).
On switching on, the display is constantly active, enabling carrying out
measuring processes. These measurements, however, cannot be saved in
moisture analyzer’s database unless a user with at least minimal access level
is logged in. The minimum access level enables carrying out measuring/drying
processes and saving them in moisture analyzer’s databases using available
function.
Below table indicates access to editing user parameters, databases
and software functions depending on active access level.
Authorization
Access level
None (guest)
No access to editing all of user parameters.
it is not possible to accept a weighing record nor to start
a process.
No access to menu <Databases>.
Operator
Access to editing parameters in submenu: <Profiles;
Readout> and settings in group of parameters
<Others>, except for <Date and Time>. The operator
can start and carry out all weighing processes, and has
access to <Export weighing database to a file> and
preview data in <Databases>. An operator can define
universal variables.
Advanced
operator
Access to editing parameters in submenu: <Profiles;
Readout>; <Working modes>; <Communication>;
<Peripherals>; <Others> except for <Date and Time>.
The advanced operator can start and carry out all
weighing processes, and can erase old data from the
<Databases>.
Administrator
Access to all user parameters, functions and editing
databases.