RADWAG MA 200.3Y Moisture Analyzer User Manual

Page 137

Advertising
background image

- 137 -

Send stable measurement result in basic measuring unit

Format: S CR LF

Accessible responses:

S_A CR LF

S_E CR LF

- command understood and in progress

- time limit exceeded while waiting for stable

measurement result

S_I CR LF

- command understood but not accessible at this

moment

S_A CR LF

MASS FRAME

- command understood and in progress

- response is mass value in basic measuring unit

Mass frame format which is moisture analyzer’s response:

1 2-3

4

5

6

7-15

16 17 18 19 20 21

S space stability

marker space character mass space

unit

CR LF

Example:
S CR LF
– command sent from a computer

S _ A CR LF - command understood and in progress

S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF - command carried out, response

is mass value in basic measuring unit.

Immediately send measurement result in basic measuring unit

Format: SI CR LF

Accessible responses:

SI_I CR LF

- command understood but not accessible at this

moment

MASS FRAME

- response is immediate with mass value in basic

weighing unit

Mass frame format which is moisture analyzer’s response:

1 2

3

4

5

6

7-15

16 17 18 19 20 21

S I space stability

marker space character mass space

unit

CR LF

Example:
S I CR LF
– command sent from a computer

S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ g _ CR LF - command carried out, immediate

response of mass value in basic weighing unit