On completing the adjustment procedure the moisture analyzer indicates

a message box:

On confirming the message box with

key, the moisture analyzer

returns to the main mode

12.2. User adjustment

User adjustment is carried out with an optional standard, which mass ranges

between 0,15 Max ÷ Max capacity. Adjustment procedure is compatible with

the external adjustment process, but before its start the software opens a

message box for entering mass of a standard used for user adjustment

procedure.

In order to start user adjustment, enter submenu < Adjustment >, and select
option: “ User adjustment”. Then follow the commands indicated on the

display.

12.3. Temperature adjustment

Temperature adjustment is carried out with a dedicated kit, that is an optional

equipment of the moisture analyzer MA 3Y series.

1. Thermometer

2. Thermometer holder and drying chamber shield

Adjustment of the drying chamber temperature is a process carried out for the

purpose of adjusting the temperature sensor assembled inside the drying