- 140 -

ES CR LF

- command not recognized (incorrect mass format)

Set high checkweighing threshold HI

Format: UH_XXXXX CR LF, where: _ - space, XXXXX – mass format

Accessible responses:

UH_OK CR LF

- command carried out

ES CR LF

- command not recognized (incorrect mass format)

Give value of low checkweighing threshold HI

Format: ODH CR LF

Accessible response: DH_MASS CR LF – command carried out

Response format:

1 2

3

4-12

13

14 15 16

17

18 19

D H

space

mass

space

unit

space

CR LF

Mass - 9 characters with right justification

Unit - 3 characters with right justification

Give value of high checkweighing threshold LO

Format: OUH CR LF

Accessible response: UH_MASS CR LF - command carried out
Response format:

1 2

3

4-12

13

14 15 16

17

18 19

U H

space

mass

space

unit

space

CR

LF

Mass - 9 characters with right justification

Unit - 3 characters with left justification



User log in

Format: LOGIN Name, Password CR LF (Name and password have to be

inserted as they are set in the moisture analyzer – small and capital letters)

Accessible responses:

LOGIN OK CR LF

- command understood, new user is logged in

LOGIN ERROR CR LF

- command understood, an error in name or

password occurred, log in failed

ES CR LF

- command not recognized (error in format)