Communication protocol, A set of commands – RADWAG MA 200.3Y Moisture Analyzer User Manual

Page 133

Advertising
background image

- 133 -

27. COMMUNICATION PROTOCOL

General information

A. A character based communication protocol moisture analyzer-terminal is

designed for establishing communication between a RADWAG instruments

and a peripheral devices via RS 232 interface.

B. It consists of commands sent from a peripheral device to a moisture

analyzer and responses from the moisture analyzer.

C. Responses are sent from the moisture analyzer on each receipt of a

command as a reaction for a specific command.

D. Commands forming the communication protocol enable obtaining data on

moisture analyzer’s status and influencing its operation, e.g.: request

measurement results from a moisture analyzer, zero indication, etc.

27.1. A set of commands

Command Command description
Z

Zero moisture analyzer

T

Tare moisture analyzer

OT

Give tare value

UT

Set tare

S

Send stable measurement result in basic measuring unit

SI

Immediately send measurement result in basic measuring

unit

SU

Send stable measurement result in current measuring unit

SUI

Immediately send measurement result in current

measuring unit

C1

Switch on continuous transmission in basic measuring unit

C0

Switch off continuous transmission in basic measuring unit

CU1

Switch on continuous transmission in current measuring

unit

CU0

Switch off continuous transmission in current measuring

unit

DH

Set low checkweighing threshold LO

UH

Set high checkweighing threshold HI

ODH

Give value of low checkweighing threshold LO

OUH

Give value of high checkweighing threshold HI

PC

Send all implemented commands

LOGIN

User log in

LOGOUT

User log out

SS

Result release

IC

Internal calibration

Advertising
This manual is related to the following products: