Minimum sample weight – RADWAG MA 200.3Y Moisture Analyzer User Manual

Page 37

Advertising
background image

- 37 -

Never – printout deactivated

First stable

– the first stable measurement result obtained after

placing a load on moisture analyzer’s weighing pan is

recorded, record of the following measurement result

takes place only on unloading the weighing pan,

returning of the mass indication below set

threshold’s value, and placing another load on

balance’s weighing pan

Last stable

– accepted is the last stable measurement recorded

before taking the load off the weighing pan. The

record takes place on taking the load off the

balance’s weighing pan, and returning of the mass

indication below set threshold’s value.

THRESHOLD – mass value obligatory for operation of automatic printout.

Set in grams.

Printout

Contains type of printout that is related to a working mode. Printout takes

place on pressing PRINT key on moisture analyzer’s overlay.

Accessible options:

STANDARD PRINTOUT

Possible declaration of printout content: HEADER, WEIGHING DATA

and FOOTER.

printout components marked as <YES> in the menu are printed on

pressing printout activating function key.

NON-STANDARD PRINTOUT

Database of printouts enables selecting one of available non-

standard printouts visible in menu <PRINTOUTS>, or designing a

unique printout which is automatically added to the databases.

Caution:

Means of designing printouts is described in point 16.5 of this user

manual.

11.7. Minimum Sample Weight

The weighing mode settings comprise a function of <Minimum sample>.

The function is enabled on specifying in the database <Database/Minimum

sample > parameters on methods for determining the minimum sample

and the values of minimal sample weight specific for a method. In standard

version, this database item is not filled in with data.

Activities related to determining minimum sample and adding data

to the database: <Databases/Minimum samples> are accessible

only for an authorized “RADWAG” personnel.

Should balance operator use the minimum sample mode, and data on

minimal sample weight are not entered to balance menu, then please

contact the local representative of “RADWAG” company.