FILTER

Each measurement signal before being displayed is electronically

processed for the purpose of obtaining corrected parameters

specific for stable measurement result, i.e. ready to read.

In a specific extend a user can influence the range of signal processing by

selecting an appropriate FILTER value. Available options:

o

very fast

o

fast

o

average / normal

o

slow

o

very slow

While setting the filtering level the user should consider the actual

operating conditions of a moisture analyzer. In case of very good

operating conditions set the filter to average/normal or fast, in case of

rough conditions set the filter to slow or very slow.

Value release

This parameter determines displaying the stability marker for a

measurement result.

There are 3 available settings of value release parameter

o

fast

o

fast + reliable /recommended/

o

reliable

Caution:

The speed of stability marker occurrence depends on applied filter value and

applied value release parameter

Autozero function

The function is to automatically monitor and correct zero

indication.

If the function is enabled, the following measurement results are

compared to each other in constant time intervals. If the results differ

less than declared AUTOZERO range, e.g. 1 division, the moisture

analyzer is automatically zeroed, and the markers of stable indication

and precise zero

are displayed.

Active AUTOZERO means, that each measurement starts from the precise

zero point. There are, however, cases where the function may disturb the

measuring process, for instance during very slow load placing or pouring

onto the moisture analyzer’s weighing pan. In such case, the zero

indication correcting system may also correct the actual indication of a

load placed on the weighing pan.

Accessible settings: NO - autozero function disabled

YES - autozero function enabled