RADWAG MA 200.3Y Moisture Analyzer User Manual

Page 44

Advertising
background image

- 44 -

chamber of a moisture analyzer. Before starting temperature adjustment

process, place the temperature calibration kit inside the drying chamber,

following below diagrams:

Step 1.

Remove from the drying chamber the

following:

•

disposable drying pan

•

drying pan handle

•

drying pan bracket

•

drying chamber shield

Step 2.

Assemble in the drying pan the

components of temperature

adjustment kit:

•

drying chamber shield

•

thermometer holder and the

thermometer

Step 3.

After assembling the kit start

temperature adjustment process.

CAUTION:

All the activities must be carried out with great caution not to damage

the measuring mechanism of the moisture analyzer.

Advertising
This manual is related to the following products: