RADWAG MA 200.3Y Moisture Analyzer User Manual
Page 143

- 143 -
Response:
OMI <CR><LF>
– available working modes are given
2_” Parts counting” <CR><LF>
number + name
4_” Dosing” <CR><LF>
12_”Checkweighing” <CR><LF>
OK <CR><LF>
– command completed
Example2
Command:
OMI <CR><LF>
– give available working modes
Response:
OMI <CR><LF>
– available working modes are given
2 <CR><LF>
(mode number)
4 <CR><LF>
12 <CR><LF>
OK <CR><LF>
– command completed
OMS - Set working mode
Command sets current working mode of a balance.
Format: OMS_n <CR><LF>
Accessible responses:
OMS_OK <CR><LF>
– command completed
OMS_E <CR><LF>
– error occurance in course of command
performance, incorrect format or lack of the
parameter
OMS_I <CR><LF>
– command understood but not accessible at this
moment
n – parameter, decimal value specifying working mode number. See OMI
command description for detailed information.
Example
Command:
OMS_13<CR><LF> – set Statistics mode
Response:
OMS_OK<CR><LF> – Statistics mode selected
OMG - Give current working mode
Command gives current working mode for a given device.
Format: OMG <CR><LF>
Accessible responses:
OMG_n_OK
<CR><LF>
– command completed, current working mode is
given
OMG_I <CR><LF>
– command understood but not accessible at this
moment