Mandatory - option enabled, the display shows a bar graph demonstrating

graphic visualization of sample mass in relation to mass value saved in the

database of products. It is a sample mass and its tolerance limits set in the

software.

Before applying this option, add a product in the database, including data on

sample mass and its checkweighing limits.

On enabling this option, the moisture analyzer’s software displays

commands on placing a sample with determined mass on the drying pan.

Unless the sample mass is within set threshold limits, the software does not

allow to start drying process.

There are two ways of using this option:

1. For

an

optional

product,

using

quick

access

keys:

<

Checkweighing thresholds> or <

Target value>.

2. In combination with a specific product saved in the <

Database of

products>.

In the 1

st

case, the target value, checkweighing thresholds and percent

tolerance of mass should be set using quick access keys.

In the 2

nd

case, the value of mass, and tolerance fields have to be specified

their corresponding fields of the <Database of products>.

SETTING TOLERANCE LIMITS FOR SAMPLE MASS:

MIN and MAX THRESHOLDS: the software accpets sample mass as

correct, if it is within set value of MIN and MAX thresholds (MIN and MAX

criteria are set in the database of products).

TOLERANCE: the software accpets sample mass as correct, if it is within

set percent tolerance limits of total sample mass (±%) - (data set in the

database of products).

Before starting drying process, a user should declare parameters of a

sample which is loaded on the drying pan: MIN and MAX, target mass and

percent tolerance of sample mass. On setting above criteria the software

allows for beginning the drying process.

1. Defining the value of Min/Max thresholds is carried out in the Database of

Products during editing a Product or by using quick access key <

Checkweighing thresholds >.

Procedure:

Press grey coloured workspace, and then <Keys> option