- 148 -

31.2. Temperature sensor cleaning

In order to ensure correct temperature measurements during testing procedure,

pay special attention to cleanness of the temperature sensor.

Any cleaning activities of the temperature sensor should be carried out with

maintaining special precautions.

REMEMBER, before initiating any maintenance or cleaning activities

of moisture analyzer’s inner parts, switch off the moisture analyzer by

pressing ON/OFF key on terminal’s overlay, and make sure the power

cord of the moisture analyzer is unplugged from mains!

Temperature sensor should cleaned using a soft fabric with addition of a mild

detergents. Do not use any abrasive agents or solvents, as they may damage

the temperature sensor.

While cleaning make sure not to touch the heating filament, as it may damage it.

If necessary it is permissible to clean filament’s shield. Clean the filament’s

shield using a soft and dry fabric only. Do not touch the filament while cleaning

the shield.