RADWAG MA 200.3Y Moisture Analyzer User Manual

Page 47

Advertising
background image

- 47 -

12.4. Drying chamber test

Parameter <Drying chamber test> enables monitoring the temperature as

indicated by the moisture analyzer’s internal thermometer.

The testing procedure is carried out using a dedicated kit, available as optional

equipment of the moisture analyzer. Means of assembling the testing kit is

described in the previous point (temperature adjustment).

In order to start the testing procedure assemble the testing kit inside the drying

chamber, and next go to menu group <Adjustment> and start the procedure by

pressing <Drying chamber test> field.

A new window is opened for

entering temperature value set for

checking the moisture analyzer.

Then enter a permissible error for

the set temperature value (the

error cannot be lower than the

default factory value). Next enter

the factory no. of the temperature

adjustment kit, and press <Run>

field. The software orders heating

the drying chamber. Control time

for a given temperature value is 8

minutes.

During the heating time the

moisture analyzer displays a

window with current data on

temperature inside the drying

chamber and heating time.

Advertising
This manual is related to the following products: