15. PROFILES

A Profile is a data pack determining:

o

functioning of working modes, e.g. weighing and drying,

o

type of data displayed during working mode operation,

o

active function keys,

o

accessible set of measuring units

o

mandatory criteria for moisture analyzer’s speed of operation and

measurement stability,

The software enables creating numerous profiles, which in practice provides:

o

each user to create their own and individual operating environment

o

moisture analyzer operation can be easily programmed by activating used

defined function keys and information on process (improving ergonomics

of operation)

15.1. Creating a profile

A default profile is named <Home>. The Administrator can create new profiles

by:

Copying an already existing profile and its modification

Creating a new profile

Copying an existing profile

Procedure:

Enter moisture analyzer’s main menu by pressing Setup key,

Enter submenu < Profiles>,

Press and hold an entry with profile name that should be copied,

A menu is opened referring to available options:

o

Edit

o

Delete

o

Copy

o

Cancel

Select option <Copy>

A new profile named <Copy name> is created, and all setting are the

same as in the base profile,

After copying change necessary data in a profile: (name, etc.)

Creating a new profile

Procedure:

Enter moisture analyzer’s main menu by pressing Setup key,

Enter submenu < Profiles>,

Press

key, which opens a message box: <Create new record?>,

Accept the message box by pressing

key. The software automatically

adds a new record and enters its editing mode.