Troubleshooting – RADWAG MA 200.3Y Moisture Analyzer User Manual

Page 150

Advertising
background image

- 150 -

31.4. Troubleshooting

Problem: no reaction to pressing main switch on/off key (no display reaction).

Probable cause:

No voltage in mains

Damaged power cord

Damaged fuse of the moisture analyzer

Moisture analyzer defect

Problem: too long pending time for drying process finish.

Probable cause:

Selected finish mode of drying process is incorrect. Carry out experimental

tests to select a proper means of finish mode.

Problem: measurements are not repeatable

Probable cause:

non-uniform sample content. Prepare the sample from larger amount of

substance.

drying time is too short. Change finish mode criteria of the testing process.

drying temperature is too high, causing sample’s oxidizing. Lower drying

temperature.

tested sample boils. Lower the drying temperature.

the temperature sensor is contaminated or defected. Clean the

temperature sensor.

the workbench on which the moisture analyzer is located is unstable.

Change the location of instrument’s operation.

- the ambient area is incompatible with the requirements (vibrations,

drafts, etc.). Change the ambient conditions for compatible ones according
to the guidelines of this user manual.

Advertising
This manual is related to the following products: