Sensors sensitivity, Proximity sensors delay, Autotest – RADWAG MA 200.3Y Moisture Analyzer User Manual

Page 128: Sensors sensitivity 23.12. proximity sensors delay

Advertising
background image

- 128 -

analyzer MA 3Y series it is not necessary to determine the drying parameters for

the products to be dried, but simply use the selected and proved drying modes.

Procedure:

Enter submenu <

Others>

Select parameter <Compatibility with MAC/MAX> which opens a window

with drop down menu,

Select one of available options:

o

YES – parameter enabled

o

NO – parameter disabled

Selecting the parameter value causes immediate returning to the menu

window.

23.11. Sensors sensitivity

The parameter is adjustable between 0 – 9, and it determines the

distance from which the IR proximity sensors should react.

As standard, the sensitivity is set within the range of 5 to 7.

Procedure:

Enter submenu <

Others>

Select parameter <Sensor sensitivity> which opens an editing window,

Select one of available options. The selection causes immediate returning

to the menu window.

23.12. Proximity sensors delay

Parameter for setting delay of proximity sensors, it is expressed in [ms].

By default the value is set to <0>.

When using option of motion recognition the user must remember about

setting <proximity sensors delay> parameter to value <500ms>

Procedure:

a) Go to <

Other> submenu.

b) Select <Proximity sensors delay> parameterand wait for a dialog

window with an on-screen keyboard to open.

c) Enter value <500>.

d) Confirm. The program returns to menu window.

23.13. Autotest

Option permanently disabled in the moisture analyzer MA 3Y series.