Introduction

1.5.3 POWER (AC Adapter)

The AC power adapter P/N 24-1404 can be used as an external power sup-

ply, whch wll run the analyzer on a contnuous bass.

When usng rechargeable NMH batteres, the AC power adapter can also

be used to charge the batteres whle nsde the analyzer. The analyzer’s

rapd-charger crcut, however, must frst be turned ON per Secton 3.14.

The rapd charger wll charge a set of depleted batteres n approxmately

2 - 3 hours.

When dsposable alkalne batteres are used, the analyzer’s battery char-

ger crcut must be OFF to prevent the batteres from overheatng. As a

precauton, the charger crcut s automatcally toggled back to ts OFF

state when the analyzer s turned OFF.

1.5.4 ∆P (Differential Pressure)

Draft s measured by connectng the probe’s draft hose to the +∆P fttng,

whle leavng the –∆P fttng open to the atmosphere.

In addton to measurng draft, the “+” and “–” ∆P fttngs can also be used

to measure the dfferental pressure between two areas by frst connectng

a hose P/N 24-1103 to the –∆P fttng, and then nsertng the open end of

ths hose nto the area beng used as the reference pressure. The analyzer’s

probe s then nserted nto the area who’s dfferental pressure s to be

measured. Refer to Secton 4.6.

1.5.5 USB (Computer Interface)

Data that has been stored n the analyzer’s memory can be downloaded to

a personal computer by connectng USB data cable P/N 104-4032 between

the USB ports of the computer and analyzer. Refer to Secton 4.14.2.

1.5.6 IrDA (Printer Interface)

Data that has been stored n the analyzer’s memory can be prnted on a

compatble IrDA wreless prnter by algnng ther IrDA communcaton

ports. Refer to Secton 4.16.

1.5.7 OPT (Option)

The opton connector s used for optonal external measurement features.