1 pca 2 dsassembly -2, 1 pca 2 disassembly – Bacharach PCA 2 24-9448 User Manual

Page 82

Advertising
background image

Instruction 24-9448

6-2

Maintenance

6.1 PCA 2 Disassembly

The followng procedure descrbes how to dsassemble the analyzer, whle

Fgures 6-1 thru 6-4 llustrate how the analyzer s put together.

Tools Requred:

Medum Phllps Screwdrver

Procedure:

1. Unplug all thermocouples from

bottom of analyzer.

2. Remove battery cover and then

remove batteres.

TIP: In Step 3, if the sensors are

not being replaced, leave the tubing

connected to each sensor’s gas cap,

being careful not to put unneces-

sary strain on the tubing during the

disassembly process.

3. Pull off sensor retaner; and

then unplug all sensors.

4. Lay analyzer face down on a padded work surface; and then usng a

medum Phllps screwdrver, remove the unt’s four rear-case screws.

5. Lft rear case from analyzer and set asde.

6. Unplug electrcal connectors J8, J9, and J14 from prnted crcut

board.

CAUTION: In Step 7, note that there is tubing connected

between the bottom hose-connector plate and the pressure sensor

on the printed circuit board. Do not put unnecessary strain on

this tubing during the disassembly process.

7. Carefully lft battery-and-pump chasss, along wth the bottom hose-

connector plate, from analyzer.

8. Lft prnted crcut board from analyzer.

Rear Case Screws

(typical 4 places)

Release and then

remove battery

cover by pushing

latch up

Figure 6-1. Removing Rear Case