Instruction 24-9448

6-11

Maintenance

6.6 Thermocouple Replacement

Usng the approprate thermocouple replacement kt lsted below, replace

the probe’s thermocouple as follows:

Thermocouple Replacement Kits:

Part Number Replaces Thermocouple in a

Probe with a Tube Length of . . .

24-8413

6 nches

24-8414

12 nches

24-8415

24 nches

24-8416

36 nches

Each kt contans a thermocouple assembly, two O-rngs, and two wre-

splce connectors.

Tools Required:
• Small Flat Blade Screwdrver

• Wre Cutter

• Wre Strpper

• Slp Jont Plers

Procedure:
1. Gan access to the thermocouple connectons by frst removng three

screws from probe handle, and then separatng the two handle peces.

2. Cut wres attached to old crmp connectors, leavng behnd as much of

the probe’s thermocouple-connector wre as possble.

Probe Handle

Screws

Figure 6-6. Probe Handle Disassembly